High pressure
Cylinder, hanger yoke, and first-stage regulator. Oxygen-cylinder pressure tracks content; nitrous oxide pressure does not until liquid is gone.
Gas pathway & delivery
MODULE 01
Follow gas from a pressurized source to the common gas outlet, then watch the circle system change with ventilation mode, fresh-gas flow, and respiratory phase. Select any machine component to turn the schematic into a guided checkout.
01 · Source to common gas outlet
Select or tab to a component. Pressure zones are color-coded; the detail pane explains mechanism, failure mode, and checkout action.
Interactive workstation map
Select any outlined component for its mechanism, failure mode, and checkout cue. Pipe colors distinguish fresh gas, inspired gas, exhaled gas, and scavenged gas.
The fail-safe valve proportionally reduces or stops N₂O flow when O₂ supply pressure falls. It senses pressure—not oxygen concentration—and therefore cannot prevent delivery of a hypoxic mixture caused by a crossed pipeline or wrong cylinder contents.
Clinical pearl
The oxygen analyzer is the only machine monitor that directly detects the inspired O₂ concentration. A functioning fail-safe is not a substitute for a calibrated analyzer.
02 · Circle system
Toggle ventilation and fresh-gas-flow strategy, then inspect inspiration and expiration. Particle speed and branching expose which valves open, which reservoir is active, and when rebreathed gas is safe.
Animated circle-system lab
Gas particles show direction rather than exact velocity. Change ventilation and flow, then inspect each respiratory phase to see which valves and pressure-relief pathways are active.
Current phase
The inspiratory unidirectional valve opens as circuit pressure or patient effort creates forward flow. The expiratory valve closes, preventing gas from taking the shorter reverse route through the expiratory limb.
Patient is the driver
The patient generates the pressure gradient. The reservoir bag is in the circuit, the APL valve controls excess pressure, and the mechanical ventilator is isolated.
Low-flow behavior
Low flow (shown as 1 L/min) conserves heat and humidity, reduces agent consumption and pollution, and increases rebreathing after CO₂ removal. Inspired concentration changes more slowly, so monitor FiO₂ and end-tidal agent closely.
Board-style safety link
The O₂ flush bypasses flowmeters and vaporizers and delivers roughly 35–75 L/min directly to the breathing system. Activating it during inspiration can cause barotrauma and can dilute inhaled anesthetic; use it deliberately, never as a routine ventilator boost.
03 · Board pattern
Cylinder, hanger yoke, and first-stage regulator. Oxygen-cylinder pressure tracks content; nitrous oxide pressure does not until liquid is gone.
Pipeline inlet, second-stage regulation, oxygen flush, and pressure-failure systems. The O₂ flush bypasses flowmeters and vaporizers.
Flow-control valves through the common gas outlet. This is the machine segment most vulnerable to leaks and misconnections.
Before every anesthetic
For education and deliberate practice—not a substitute for supervision, patient-specific judgment, device instructions, or local emergency protocols.