Gas pathway & delivery

MODULE 01

Anesthesia machine & breathing circuits

Follow gas from a pressurized source to the common gas outlet, then watch the circle system change with ventilation mode, fresh-gas flow, and respiratory phase. Select any machine component to turn the schematic into a guided checkout.

Supply pressureFlow controlAgent deliveryRebreathing circuit

01 · Source to common gas outlet

Map the workstation

Select or tab to a component. Pressure zones are color-coded; the detail pane explains mechanism, failure mode, and checkout action.

Interactive workstation map

Follow gas from source to patient

Select any outlined component for its mechanism, failure mode, and checkout cue. Pipe colors distinguish fresh gas, inspired gas, exhaled gas, and scavenged gas.

Fresh gasInspired gasExhaled gasScavenging
Gas delivery systemCircle breathing systemCGOoutletto scavengerO₂~2000 psigN₂O~745 psigAIR~2000 psigCylinders + pressure regulatorsFail-safesenses O₂ pressureN₂OAIRO₂FlowmetersO₂ downstreamSEVOVariable-bypass vaporizerPatient + Y-pieceInspiratory valveExpiratory valveCO₂ absorber · heat + waterAPL valve2–3 L reservoir bagspillvalveVentilator bellows
Intermediate-pressure systemSelected component

Oxygen pressure–failure protection device

The fail-safe valve proportionally reduces or stops N₂O flow when O₂ supply pressure falls. It senses pressure—not oxygen concentration—and therefore cannot prevent delivery of a hypoxic mixture caused by a crossed pipeline or wrong cylinder contents.

Clinical pearl

The oxygen analyzer is the only machine monitor that directly detects the inspired O₂ concentration. A functioning fail-safe is not a substitute for a calibrated analyzer.

The oxygen pressure–failure device reacts to supply pressure, not gas identity. A crossed pipeline can be fully pressurized and still deliver the wrong gas. The calibrated inspired-oxygen analyzer is the final defense against a hypoxic mixture. When a volatile agent is introduced, its output and clinical effect are separate questions—review alongside the vaporizer pathway.

02 · Circle system

Run the gas-flow simulation

Toggle ventilation and fresh-gas-flow strategy, then inspect inspiration and expiration. Particle speed and branching expose which valves open, which reservoir is active, and when rebreathed gas is safe.

Animated circle-system lab

Change the driver and fresh-gas flow

Gas particles show direction rather than exact velocity. Change ventilation and flow, then inspect each respiratory phase to see which valves and pressure-relief pathways are active.

Ventilation mode
Fresh-gas flow
FRESH GAS1 L/minPATIENTnegative-pressure effortINSPIRATORYOPENEXPIRATORYCLOSEDAPL · OPENRESERVOIR BAGCO₂ ABSORBERexothermic · produces H₂OVENTILATOR + SPILL VALVEfreshinspiredexhaled
Inspect respiratory phase
Inspiratory valveOpen
Expiratory valveClosed
Bag + APL branchActive
Ventilator branchIsolated

Current phase

Gas moves through the inspiratory limb toward the patient

The inspiratory unidirectional valve opens as circuit pressure or patient effort creates forward flow. The expiratory valve closes, preventing gas from taking the shorter reverse route through the expiratory limb.

Patient is the driver

The patient generates the pressure gradient. The reservoir bag is in the circuit, the APL valve controls excess pressure, and the mechanical ventilator is isolated.

Low-flow behavior

Low flow (shown as 1 L/min) conserves heat and humidity, reduces agent consumption and pollution, and increases rebreathing after CO₂ removal. Inspired concentration changes more slowly, so monitor FiO₂ and end-tidal agent closely.

Board-style safety link

The O₂ flush bypasses flowmeters and vaporizers and delivers roughly 35–75 L/min directly to the breathing system. Activating it during inspiration can cause barotrauma and can dilute inhaled anesthetic; use it deliberately, never as a routine ventilator boost.

03 · Board pattern

Three pressure zones, three failure signatures

up to 2,000 psig

High pressure

Cylinder, hanger yoke, and first-stage regulator. Oxygen-cylinder pressure tracks content; nitrous oxide pressure does not until liquid is gone.

~45–55 psig

Intermediate pressure

Pipeline inlet, second-stage regulation, oxygen flush, and pressure-failure systems. The O₂ flush bypasses flowmeters and vaporizers.

near atmospheric

Low pressure

Flow-control valves through the common gas outlet. This is the machine segment most vulnerable to leaks and misconnections.

Before every anesthetic

Checkout is a functional test, not a visual ritual

1Verify backup ventilation and suction
2Confirm O₂ supply and analyzer calibration
3Leak-test machine and breathing circuit
4Test ventilator, alarms, scavenging, and monitors

For education and deliberate practice—not a substitute for supervision, patient-specific judgment, device instructions, or local emergency protocols.