Mechanics made visible

MODULE 04

Ventilator modes & waveform lab

Adjust the ventilator, change respiratory mechanics, and create common intraoperative faults. Pressure, flow, volume, and capnography redraw continuously so the pattern and its physiology stay connected.

VCV / PCV / PSVCompliance & resistanceFour synchronized curvesFive scenarios

Live simulation

Change one variable, read all four curves

Start with normal mechanics. Then isolate resistance, compliance, disconnection, and patient effort. Use the calculated pressure relationships to explain what changed before revealing the scenario notes.

Interactive physiology lab

Ventilator waveform laboratory

Change ventilator targets or respiratory mechanics, then compare pressure, flow, volume, and capnography in real time.

Ventilation mode
Clinical scenario

Normal mechanics: Reference waveforms with the selected lung mechanics.

Animated ventilator pressure, flow, volume, and capnography waveforms.

Volume is controlled; pressure is the dependent variable. Inspiratory flow is positive, expiratory flow is negative, and volume returns toward baseline before the next breath. Capnography has a rapid upstroke and near-horizontal plateau.

Estimated Ppeak

20cmH\u2082O

Estimated Pplat

15cmH\u2082O

Estimated V\u1d1b

500mL

Time constant

0.50s

Ventilator and lung settings

12 /min
630
500 mL
250800
5 cmH\u2082O
016
50 mL/cmH\u2082O
20100
10 cmH\u2082O/L/s
540
In volume control with constant inspiratory flow, peak pressure reflects resistive plus elastic pressure; plateau pressure approximates elastic pressure after flow stops. A rising peak with a stable plateau suggests increased airway resistance. A rise in both peak and plateau suggests decreased respiratory-system compliance. Always confirm the tracing and examine the patient and circuit.

VCV

Volume targeted

Delivered volume is set; pressure varies with resistance and compliance.

PCV

Pressure targeted

Inspiratory pressure is set; tidal volume varies with mechanics and time.

PSV

Patient triggered

Each spontaneous breath receives pressure assistance; apnea requires a backup plan.

Capnogram

Ventilation + perfusion

Shape, slope, baseline, and trend often localize a fault before a single number does.

For education and deliberate practice—not a substitute for supervision, patient-specific judgment, device instructions, or local emergency protocols.